Sensor cover
Display
Sample/settings button

Memory recall button
105mm
58mm

8mm
33mm
Test cassette
Thumbwheel
Battery cover
Lancet cover

Sensor
Penetration depth indicator

8mm
33mm
19.5mm

Penetration depth selector
Lancet holder
Lancet
Lancet ejector
Technical data
Meter Name | Betachek C50 |
Intended use | Self testing of blood glucose. Not suitable for multi patient use |
Integrated Lancing Device | Single patient use only 10 penetration depths with eject mechanism Lancet: Betachek lancets, Softclix lancets |
Test Time | Approx. 5 seconds |
Measuring method | Reflex photometry |
Compliance | CE, TGA, EN ISO15197:2015 |
Assay method | Glucose Oxidase |
Blood sample | Fresh capillary blood only |
Sample size | Minimum: 0.3 μL Maximum: 5μL |
Units of measure | mmol/l or mg/dl – factory set. |
Power supply | 1 battery (type CR2032) |
Battery life | 1000 tests or 1 year |
Memory | 500 results with time and date, averages for 7,14,30 and 90 days |
Measurement range | 20-600mg/dL (1.1 – 33.3 mmol/L) |
Auto power off | 120 seconds |
Temperature | During testing 10 ̊C to 40 ̊C (50 ̊F to 104 ̊F) During storage without battery and test cassette: -25 ̊C to + 70 ̊C (-13°F to + 158°F) During storage with battery and without test cassette: -20 ̊C to 50 ̊C (-4 ̊F to +122 ̊F) During storage with battery and test cassette: +2 ̊C to +30 ̊C (+35.6 ̊F to 86 ̊F) |
Atmospheric Humidity | During testing 15-85% relative humidity |
Altitude | Sea level to 4000m |
Meter Dimensions | 105 x 58 x 19.5mm |
Cassette Dimensions | 22.2 x 13 x 48.2mm |
Cassette expiry date | 24 months from date of manufacture 3 months from date of opening |
Cassette interface | RFID to connect with meter |
Meter Interface | Bluetooth - smartphone RFID - cassette |
Weight | Meter 60g; Meter, battery, cassette 70g |
Model/Serial Number | Located on meter label |
Safety Class | III |
LED | Class 1 |